Abstract: The Soft X-ray Spectrometer instrument on the Astro-H observatory contains a 6×6 array of x-ray microcalorimeters that is cooled to 50 mK by an adiabatic demagnetization refrigerator (ADR). The ADR consists of three stages in order to provide stable detector cooling using either a 1.2 K superfluid helium bath or a 4.5 K Joule-Thomson (JT) cryocooler as its heat sink. When liquid helium is present, two of the ADR’s stages are used to single-shot cool the detectors while rejecting heat to the h…